25 April 2012

Differential vs. Single-ended Measurements

Apogee SO series oxygen sensor
A couple of years ago, while talking to a customer who had a limited number of input channels on his datalogger, I suggested measuring an oxygen sensor using a single-ended channel instead of a differential channel. He questioned the suggestion, thinking that a differential signal can only be measured using a differential channel. I used this opportunity to educate him about the difference between differential and single-ended measurements. An oxygen sensor can indeed be measured using a single-ended channel, by wiring the low output of the sensor to ground rather than the low input on the differential channel, but there are some trade-offs. This blog post outlines the trade-offs between single-ended and differential measurements, and describes when differential is necessary and when single-ended can be used.

A differential voltage is “floating”, meaning that it has no reference to ground. The measurement is taken as the voltage difference between the two wires. The main benefit of a differential measurement is noise rejection, because the noise is added to both wires and can then be filtered out by the common mode rejection of the data acquisition system. Differential measurements should be used if the sensor is in a noisy environment or for sensors with output voltages susceptible to noise interference. For example, we recommend that the thermopile output from Apogee SI-100 series infrared radiometers should always be measured differentially because the small voltages are susceptible to noise.

A single-ended measurement is taken as the voltage difference between a wire and ground. The noise is only on the positive wire, and as a result, it is still measured along with the output voltage from the sensor. Some sensors, for example amplified versions of Apogee SP and SQ series sensors, only have a single output and must be wired into a single-ended channel. A sensor with a differential output can be wired for single-ended by wiring the low side to ground. This is usually done to reduce the number of channels needed to measure the sensors. It should be noted that some sensors (for example, the thermopile output from Apogee SI-100 series infrared radiometers) can output a negative voltage which means that the data acquisition system needs to be able to measure negative voltages.

It should also be noted that taking a single-ended vs. differential measurement might also be based on the data acquisition system or the cabling. Shielded twisted-pair cable is also very effective at reducing noise in the signal other types of cable might not have the same effect. The data acquisition system is also important to consider if it doesn’t have a very uniform ground the signal could be biased. In most cases, modern data acquisition systems have improved in this area and it isn’t as much of a concern.

It should be stated that differential measurements should always be used if there are enough available datalogger channels or if the data acquisition system cannot measure negative voltages. Table 1 (below) shows which measurement method should be used with specific Apogee Instruments sensors. Again, it should be remembered that any in the differential measurement column can be measured single-ended if the conditions warrant it.

Table 1: Apogee Instruments Sensors Output
Differential Single-ended
SI-100 series - thermopile
(single-ended measurement strongly discouraged)
SI-100 series - sensor body temperature
SP-100 series SP-200 series
SO-100/200 series SF-110
SQ-100/300 series SQ-200 series
SU-100 ST-100



 

 


Skif Smith
Electrical Engineer

No comments:

Post a Comment